Martindale MARHPAT600
PAT Tester Hand Held
Product Overview
Manual and Automatic Modes
In automatic mode, the HPAT600 will routinely run through a selected test sequence without any further user interaction unless a failure is encountered.
In manual mode, a connection diagram for the selected test type is shown on the display and each step of a test requires user confirmation by a press of the OK button.
The user can select either the standard 500V insulation test or a 250V option which is required for testing appliances with surge protected circuits. The pass levels can also be adjusted which is particularly useful for testing appliances with long power cords or long extension leads.
Stores Data For 200 Appliances
The HPAT600 stores test data for up to 200 appliances which can be easily reviewed on the instrument if you need to check a result on site. It also has a simulated load test to check the load and ensure that the appliance is switched on with a good fuse.
Product Features
Tests both 240V and 110V appliancesProduct Specs
- Test Current: 200mA DC
- Test Voltage: 4V DC
- Range: 0 - 19.99Ω
- Resolution: 0.01Ω
- Default Pass Value: 0.1Ω
- User Selectable Pass Value: 0.1Ω to 0.9Ω in 0.1Ω steps
Insulation Resistance
- Test Voltage:
500V: >500V DC at 1mA
250V: >250V DC at 1mA - Default Pass Value: Class I - 1MΩ, Class II - 2MΩ
- User Selectable Pass Value: 0.3MΩ to 1MΩ in 0.1Ω steps, or 1MΩ to 10MΩ in 1MΩ steps
IEC & Extension Lead Test
- Earth Continuity: 200mA DC
- Insulation Test: 500V DC
- Wiring Test: Live, Neutral Continuity& Polarity
- Earth Bond Default Pass Value: 0.1Ω
- Earth Bond User Selectable Pass Value: 0.1Ω to 0.9Ω in 0.1Ω steps
- Insulation Resistance Default Pass Value: 1MΩ
- Insulation Resistance User Selectable Pass Value: 0.3MΩ to 1MΩ in 0.1Ω steps, or 1MΩ to 10MΩ in 1MΩ steps
Fuse Test
- Test Voltage: 4V DC
- Test Current: 200mA DC
- Pass Level: >2mA